User Guide
True RMS Multimeter with
IR Thermometer
Extech EX210T
Przeglądanie stron 0
1 2 3 4 5 6 ... 11 12

Podsumowanie treści

Strona 1 - IR Thermometer

User Guide True RMS Multimeter with IR Thermometer Extech EX210T

Strona 2 - Introduction

EX210T-EU-EN v1.2 08/13 10Maintenance WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the ba

Strona 3 - CAUTIONS

EX210T-EU-EN v1.2 08/13 11Specifications Function Range Resolution Accuracy DC Voltage 200 mV 0.1 mV (0.8% reading + 6 digits) 2V 0.001V (0.5

Strona 4

EX210T-EU-EN v1.2 08/13 12Function Range Resolution Accuracy Resistance 200  0.1  (0.8% reading + 5 digits) 2 k 0.001 k (0.8% reading + 2 d

Strona 5 - Symbols and Enunciators

EX210T-EU-EN v1.2 08/13 2Introduction Thank you for selecting the Extech EX210T True RMS Autoranging Multimeter. This meter measures AC/DC Voltage,

Strona 6 - Operating Instructions

EX210T-EU-EN v1.2 08/13 3PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY OVERVOLTAGE CATEGORY I Equipment of OVERVOLTAGE CATEGORY I is equipment for c

Strona 7

EX210T-EU-EN v1.2 08/13 4SAFETY INSTRUCTIONS This meter has been designed for safe use, but must be operated with caution. The rules listed below mu

Strona 8

EX210T-EU-EN v1.2 08/13 5Controls and Jacks 1. IR Thermometer 2. Laser Pointer 3. 2000 count LCD 4. MAX button 5. MODE button 6. FUNCTION switch

Strona 9

EX210T-EU-EN v1.2 08/13 6Operating Instructions WARNING: Risk of electrocution. High-voltage circuits, both AC and DC, are extremely dangerous and s

Strona 10 - Maintenance

EX210T-EU-EN v1.2 08/13 7RESISTANCE MEASUREMENTS WARNING: To avoid electric shock, disconnect power to the unit under test and discharge all capacit

Strona 11 - Specifications

EX210T-EU-EN v1.2 08/13 8NON-CONTACT INFRARED TEMPERATURE MEASUREMENTS 1. Rotate the function switch to the IR position. The upper right four digit

Strona 12 - ISO‐9001Certified

EX210T-EU-EN v1.2 08/13 9AUTORANGING/MANUAL RANGE SELECTION When the meter is first switched on, it automatically enters the AutoRanging mode. This